Reference Manual
300550en, Rev AA
December 2010
4-3
Rosemount 2240S
Section 4. 2240S Installation
4.3 MECHANICAL
INSTALLATION
4.3.1
Mounting on top of a Temperature Sensor/WLS
1. Remove the plug.
2. Unscrew the four screws and remove
the cover.
3. Remove the plug that protects the cable
entry at the bottom of the 2240S
transmitter housing.
4. Attach the 2240S transmitter on top of
the temperature sensor tube.
5. Remove the cover and run the sensor
wires into the terminal compartment.
6. Tighten the nut on the 2240S
transmitter by hand.
7. Proceed with electrical installation of
temperature elements and water level
sensor.
Plug
Cover screw
(x4)
Cover
Sensor wires
Sensor wires
Sensor wires
Sensor wires
Nut