
R&S FSH
Distance-to-Fault Mode (R&S FSH-K41)
Configuring Cable and Antenna Tests
Operating Manual 1309.6275.12 - 14
161
Adjusting the vertical axis automatically
The R&S FSH provides an automatic scaling routine that scales the vertical axis in a
way that the results fit ideally on the display. The R&S FSH does this by determining
the minimum and maximum trace values and scaling the vertical axis according to
these values.
M
Press the AMPT key.
M
Press the "Range" softkey.
M
Select the "Auto Scale" menu item.
The R&S FSH performs automatic adjustment of the vertical axis.
Setting the reference position
The reference position defines the position of the reference line in the diagram. The
reference position is a linear value between 0 and 10. Each value represents one
horizontal grid line of the diagram. 0 corresponds to the top grid line and 10
corresponds to the bottom grid line.
When you change the reference position, the R&S FSH also shifts the position of the
trace by the magnitude of the reference position change. It has no effect on the
reference value itself.
M
Press the AMPT key.
M
Press the " Ref Pos" softkey.
M
Enter the reference position you want.
The R&S FSH moves the trace accordingly.
5.2.3.2 Setting the Attenuation
The R&S FSH provides functions to attenuate the signal at both the tracking generator
output and the RF input.
M
Press the AMPT key.
M
Press the "TG Output Attenuation" softkey.
M
Enter the attenuation of the signal at the tracking generator output.
You can attenuate the signal from 0 dBm to -40 dBm.
An input of 10 dB, for example, corresponds to an attenuation of -10 dBm.
M
Press the "RF Att/Amp/Imp" softkey.
M
Select the "Manual:" menu item or one of the automated attenuation modes.
M
Enter the attenuation of the signal at the RF input.
You can set an RF attenuation in the range from 0 dB to 40 dB in 5 dB steps.
Содержание FSH4
Страница 1: ...R S FSH4 8 Spectrum Analyzer Operating Manual 1309 6275 12 14 Test Measurement Operating Manual ...
Страница 19: ......
Страница 245: ......