R&S FSH
Network Analyzer Mode
Performing Vector Measurements (R&S FSH-K42)
Operating Manual 1309.6275.12 - 14
133
4.3.1 Measuring the Transmission
The example included here is based on a transmission measurement of a two-port
filter. The filter works in the frequency from of 1920 MHz to 1980 MHz.
Test setup
M
Connect the input of the DUT to the tracking generator output (port 2).
M
Connect the output of the DUT to the RF input (port 1).
Preset the R&S FSH
Before starting the measurement procedure, preset the R&S FSH to restore the default
configuration and connect the filter between the measurement ports.
M
Press the PRESET key.
M
Connect the DUT.
Start vector measurements.
M
Press the MEAS key.
M
Press the "Meas Mode" softkey.
M
Select the "Vector" menu item.
Select the type of transmission measurement
M
Press the "Result Display" softkey.
M
Select the "Transmission Fwd (Port 1 2)" menu item for transmission
measurements in forward direction or "Transmission Rev (Port 1 2)" menu item
for transmission measurements in reverse direction.
Define the frequency parameters
Before calibrating the measurement, you should define the frequency parameters to
avoid interpolation of the results.
M
Press the FREQ key.
The R&S FSH opens an input field to set the center frequency.
M
Enter a frequency of 1950 MHz.
M
Press the SPAN key.
The R&S FSH opens an input field to define the span.
M
Enter a span of 130 MHz to get a high resolution of the results.