R&S FSH
Working with Power Sensors
Using a Directional Power Sensor
Operating Manual 1309.6275.12 - 14
121
If high powers are applied that exceed the maximum input level of the R&S FSH-Z14
or R&S FSH-Z44, a directional coupler or an attenuator has to be connected ahead of
the power sensor. In such cases, the coupling attenuation of the directional coupler or
the attenuation value of the attenuator are to be entered as positive dB values (see
above) into the R&S FSH to ensure true measured power readout. In both cases, a
termination or an attenuator of sufficient power-handling capacity has to be connected
to the power sensor at the load end. The matching readout is irrelevant in such case
since it is likewise corrected by taking into account the attenuation value of the
termination or attenuator.