4.5.2.1 Performing Transmission Measurements
Calibrate the measurement for transmission measurements.
M
Press the "Calibration" softkey.
M
Select the "Transmission Rev
(Port 2 1)" menu item.
M
Perform calibration. For more
information see "
M
Reconnect the DUT.
The R&S FSH shows the
characteristics of the DUT
(magnitude and phase) as
numerical values.
You can change the measurement configuration (e.g. the sweep time) or format
(see "
Selecting the Measurement Format
") without affecting the accuracy of the
measurement.
4.5.2.2 Performing Reflection Measurements
Calibrate the measurement for reflection measurements.
M
Press the "Calibration" softkey.
M
Select the "Reflection Port 1" menu item.
M
Perform calibration. For more information see "
M
Reconnect the DUT.
The R&S FSH shows the characteristics of the DUT (magnitude and phase) as
numerical values.
You can change the measurement configuration (e.g. the sweep time) or format
(see "
Selecting the Measurement Format
") without affecting the accuracy of the
measurement.
Note that for calibration to remain valid, frequency parameters, bandwidth and
attenuation have to remain the same.