Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter •
6
85
6.3
Storing linearity test information
Observe the applicable regulations and directives of the
responsible regulatory agencies when performing linearity
tests.
Entering the lot number of the
linearity test
The following description assumes that the meter is
powered on and the
Main Menu
is displayed.
1
Touch
to open the
Main Menu 2
screen.
2
Touch
Linearity Lots
to open the related menu.
3
Touch
Add
to enter a new lot number.
Main Menu
Review Results
Control Test
Patient Test
12:48 pm
09/18/13
Main Menu 2
Maintenance
Strip Lots
Profi ciency
Control Lots
Beeper
Linearity
Diagnostics
Linearity Lots
Admin.
12:48 pm
09/18/13
Linearity Lots
Add
* Lin.
12/31/13
777732
03/15/14
777723
Lin.
Type Exp. Date
Lot Num.
12:48 pm
09/18/13
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...