9
• Initial Startup
118
9.7
Display options and optional tests
Use this menu to select the unit of measure for test
results and enable or disable optional tests.
1
Touch the desired option to enable it:
■
Result Units
– mg/dL
– mmol/L
■
Enable Tests
–
STAT
: STAT tests may be performed. When this
option is enabled, you must enter the number of
STAT tests allowed (after confirming the settings).
–
Linearity
is displayed in the
Main Menu 2
screen
–
Proficiency
is displayed in the
Main Menu 2
screen
–
Maintenance
is displayed in the
Main Menu 2
screen
2
Touch
to
save
the
settings.
If
the
STAT
option
is enabled, enter the number of STAT tests allowed,
or touch
to exit this menu without saving any
changes.
The display returns in both cases to the
Setup Menu
(after entering the number of STAT tests allowed).
Test Disp. Options
mg/dL
mmol/L
Enable Tests:
Stat
Linearity
Profi ciency
Maintenance
Display Strip Limitations
Warning
Result Units
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
STAT Test Allowed
8
12:48 pm
09/18/13
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...