10
• Maintenance and Care
134
How to clean/disinfect
1
Remove the meter from the base unit prior to
cleaning/disinfecting.
2
Power off the meter.
3
Place the meter on a level surface.
4
Gently wipe over the surfaces (touchscreen,
housing) with a soft, lint-free cloth slightly
dampened
(not wet)
.
If using commercially available pre-moistened cleaning/
disinfection wipes,
squeeze off excess solution
or blot
on a dry paper towel to remove any excess solution
before wiping the surface of the meter.
5
Carefully wipe over the test strip port area, making
sure that no liquid enters the test strip port.
If spraying the meter,
place it on a flat surface or
table. Wear gloves. Be aware that the meter may
become slippery when wet. Do not let the meter
drop! Do not spray while the meter is docked in the
base unit!
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...