Glucose Control Testing •
4
57
4
Glucose Control Testing
4.1
Information regarding glucose control tests
Observe the applicable regulations and directives of the
responsible regulatory agencies when performing glu-
cose control tests. See also safety message “Allergy or
injury caused by reagents and other working solutions”
on page 18.
Accurately testing known levels of glucose ensures that
the system and your technique used in testing give accu-
rate results on patient tests. Glucose control solutions
have defined (known) values. The results for these solu-
tions must first fall within a certain acceptable range in
order to allow valid patient testing.
The system may be configured to require in-range glu-
cose control testing before patient testing is allowed. This
is called
QC Lockout
, and the system actually prevents
patient and proficiency testing when the control results
are not within the accepted range.
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...