Initial Startup •
9
123
9.10 Value ranges (normal, critical, reportable)
This menu allows you to set limit values for results as
Normal
,
Critical,
or
Reportable
. Results outside these
limits will be flagged to alert you of this event. The
Reportable Range
allows the system administrator to set
an institution defined range for reporting patient results.
Values for range limits are entered one directly after the
other in the following order:
■
Normal (low/high limit value)
■
Critical (low/high limit value)
■
Reportable (low/high limit value)
1
Use the keypad to enter the first value.
2
Touch
to enter the following value, or touch
to return to the previous view.
3
After entering the last value, touch
to save all
the previous settings.
1
2
3
4
5
6
7
8
9
0
Report. Range Lo
0.6
mmol/L
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Normal Range Lo
3.9
mmol/L
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Critical Range Lo
2.2
mmol/L
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Report. Range Hi
33.3
mmol/L
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Normal Range Hi
11.1
mmol/L
12:48 pm
09/18/13
1
2
3
4
5
6
7
8
9
0
Critical Range Hi
16.7
mmol/L
12:48 pm
09/18/13
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...