Glucose Control Testing •
4
67
Results screen
The hourglass icon indicates the test is running. When
the test is completed and the result is ready, the meter
beeps again.
Depending on configuration, the result is displayed either
as a value or only as a qualitative result
Pass
or
Fail
. Your
system may be configured to disallow further testing until
all the required glucose control levels are successfully
run (QC Lockout).
When the results are displayed as a value, the results
screen contains a button that changes its name based on
the result (
Range
or
Out of … Range
). Touch this button
to display the minimum and maximum target values for
the control levels.
You can add comments to a test result (as with blood
glucose tests) (see page 54).
If you do not wish to add a comment to the test result,
touch
to continue to the next level of the control
test, if necessary, or to return to the
Main Menu
.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives.
Control Test
PASS
Control L2 (Hi) 123456
Date 09/18/13 12:48 pm
12:48 pm
09/18/13
Control Test
Range
Control L2 (Hi) 123456
Date 09/18/13 12:48 pm
17.1
mmol/L
09/18/13
12:48 pm
Control Test
Control L2 (Hi) 123456
Date 09/18/13 12:48 pm
FAIL
12:48 pm
09/18/13
Test results are also saved when the meter is powered
off in the results screen or when it automatically
powers off.
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...