Linearity Testing •
7
95
Applying a linearity test sample
Once the meter has checked the test strip, the hourglass
icon disappears and you are prompted to apply linearity
solution.
1
Wait until the flashing drop appears in the display
before applying the solution. The meter beeps.
2
Apply a drop of the linearity solution to the
front
edge
of the test strip. Do
not
apply the solution to
the top of the strip.
The linearity solution is pulled into the test strip
area by capillary action.
Once sufficient linearity solution has been detected, the
meter beeps and the measurement begins.
Linearity Test
Linearity 12345678 L1
Strip Lot 545794
12:48 pm
09/18/13
When applying the linearity solution, position the
meter so that the test strip port is always higher or on
the same level as the linearity solution. This prevents
any excess solution from flowing down the strip and
entering the meter.
Содержание ACCU-CHEK Inform II
Страница 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Страница 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Страница 6: ...6 This page intentionally left blank ...
Страница 12: ...12 This page intentionally left blank ...
Страница 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Страница 160: ...12 General Product Information 160 This page intentionally left blank ...
Страница 174: ...A Appendix 174 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 GS1 DataBar Limited PDF417 DataMatrix QR Code ...
Страница 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Страница 210: ...Index 210 This page intentionally left blank ...
Страница 211: ...211 ...