background image

If you see the letter “E” followed by a number, or if your Meter beeps twice, an error has
occurred. The following chart gives you a list of a few possible errors and describes what they mean.

What You See

What it Means

What You Should Do

Low battery

Replace the batteries as soon as possible. Bayer HealthCare offers free 
replacement batteries for Ascensia customers. Call Customer Service at
1-800-348-8100 (24 hours a day).

E 1

Temperature out 

Test only when the meter temperature is between 

of range 

50°F and 104°F (10°C and 40°C). 

E 2 

Test Strip underfill

Remove the Test Strip and repeat the test with a new Test Strip.

E 3

Used Test Strip

Replace Test Strip with a new, unused Test Strip. Wait until you see the
flashing blood drop before you add blood sample.

E 4

Test Strip not 

Remove the Test Strip and insert it properly. 

inserted correctly

E 5 and above

All other errors

When you see the letter E followed by a number that is not listed here,
remove Test Strip and retest with a new Test Strip. If you continue to 
experience difficulty, call Customer Service for assistance.

LO

Test result is below 

Repeat test. If LO appears again, contact your Physician or Healthcare 

10 mg/dL (0.6 mmol/L)

Professional immediately. CAUTION: Glucose levels below 50 mg/dL 
(2.8 mmol/L) may indicate a potentially serious medical condition.

HI

Test result is above 

Wash your hands and test site. Repeat the test. If the test result is still 

600 mg/dL 

“HI,” contact your Physician or Healthcare Professional immediately.

(33.3 mmol/L)

CAUTION: Glucose levels above 250 mg/dL (13.9 mmol/L) may 
indicate a potentially serious medical condition.

Solving Problems

© 2003 Bayer HealthCare LLC

Made and Printed in U.S.A.

50184256    

Rev. 9/03

Note: See “Solving Problems” section of your 

Ascensia

®

C

ONTOUR

™ User Guide for further 

information and a complete list of errors.