69
4-5
Printing
4-5-1 Printing Procedure
Print out the data transferring by IR rays from the instrument.
Measurement mode and data transmittable distance
Measurement mode
Distance between
main unit and printer
AUTO, AUTO2
CONT FORMAT “ALL”
CONT FORMAT “LAST5”
Sending stored data
“SELECT”
Sending stored data
“LAST10
・
30”, “ALL”
Less than 50cm
○
○
50cm or more
○
×
Distance from instrument and
the printer
Data receiving
window
If measurement mode is “CONT”, in case setting <OUTPUT DATA> in
[OUTPUT] - [CONT FORMAT] with [SETUP] to “ALL”, an error will occur if
data is transmitted while the main unit and the printer are separated by 50 cm
or more. Be sure to transmit the distance between the main unit and the
printer closed to less than 50cm. If <OUTPUT DATA> is set to “LAST5”, it can
be transmitted even if it is more than 50cm apart.
In the case of “transmitting stored data”, if <OUTPUT NUMBER> in
[MEMORY] with [SETUP] is set to “LAST10”, “LAST30”, or “ALL”, an error
occurs when the data is transmitted from 50cm or more apart. Be sure to
transmit the distance between the main unit and the printer closed to less
than 50cm. If <OUTPUT DATA> is set to “SELECT”, it can be transmitted
even if it is more than 50cm apart.
Содержание Retinomax K+Screeen
Страница 1: ...Hand Held Auto Refractometer Instructions November 2018 D054 E...
Страница 22: ...21 Manufacture date Station Rear of the station...
Страница 23: ...22 Printer Bottom Battery part Manufacture date...
Страница 24: ...23 Battery pack...
Страница 65: ...64 5 When shifting to QUICK mode by AUTO QUICK mode AQ is printed on the measured data Printing example...
Страница 112: ......