106 

 

8-3 

Printer 

Phenomenon 

Check point 

Cause and action 

The printer connected to 
the station does not turn 
on when the printer power 
switch is flipped on. 

Is the printer correctly connected to 
the station? 

Correct the connection. 
See “3-4-2 Automatic Charging of 
the Printer”. 

Is the contact (5 places) dirty? 

If touching the contacts (5 places) or 
getting dirty, wipe it off with a dry 
cloth. 

Is the station in a power-off state? 

The printer will not operate if the 
station is not turned on. Turn on the 
station. 

The battery pack is set in 
the printer, but the power 
lamp does not light when 
the power switch is set to 
on. 

Is the battery pack in the correct 
orientation? 

Check that the battery pack is 
correctly installed. 
See “3-2 Installing and Removing 
the Battery Pack”. 

Is the lamp disappeared after 5 sec 
when it blinked with orange? 

Recharge the battery pack. 
See “3-4 Charging the Battery 
Pack”. 

When the printer power 
lamp flashed pink color 

Is the paper holder cover open? 

Close the paper holder cover. 

Is the print roll set correctly? 

Set the print roll correctly. 
See “7-2 Replacing the Print Roll”. 

Was the measurement data printed 
when it transmitting? 

Or, was it transferred to computer? 

Confront instrument transmission 
window with the printer’s receiving 
window and transfer again. 

The printer does not 
receive measurement data 
sent from the main unit. 

Are there any obstacles between 
the main unit and printer? 

Remove the obstacles, or move to a 
location free of obstacles. 

Is the printer too far away from the 
main unit? 

Data can be transmitted to a 
location within 6 meters from the 
printer and with an angle of less 
than 45 degrees between the front 
side of the main unit and the light 
receiving window of the printer. 
And, if the mode were ”CONT” or 
“Transmitting from Memory 
function”, it cannot transfer over 
range of 50cm. 
See “4-5-1 Printing Procedure”. 

Is the ID number of the printer 
same as that of the main unit? 

Match the ID numbers by checking 
the OUTPUT screen and the DIP 
switch in the printer. 
See “4-5-4 Using Multiple Printers”. 

Characters are crowded 
on the print paper. 
The paper does not come 
out properly. 

Is the print roll set correctly? 

Set the print roll correctly. 
See “7-2 Replacing the Print Roll”. 

Are there red lines on the edges of 
the print paper? 

Replace the print roll. 
See “7-2 Replacing the Print Roll”. 

Cannot connect to 
external devices. 

Is the communication setting of this 
unit and the external device 
correct? 

Make sure the setting of each 
devices. See “6. Connecting to 
External Instruments”. 

Cannot send the data to 
RV-II. 

Are there anything to interfere the 
connection between devices? 
(The case that connecting with IR to 
the RV-II.) 

Make sure there are nothing to 
interfere the communication. 

Содержание Retinomax K+Screeen

Страница 1: ...Hand Held Auto Refractometer Instructions November 2018 D054 E...

Страница 2: ...lectrotechnical Commission IEC Before using the instrument be sure to read the instructions carefully and fully understand the operation procedures and safety instructions to ensure correct usage Also...

Страница 3: ...t 30 3 2 2 Printer 31 3 3 Setting Up the Instrument 31 3 4 Charging the Battery Pack 32 3 4 1 Automatic Charging of the Main Unit 33 3 4 2 Automatic Charging of the Printer 34 3 4 3 Charging a Spare B...

Страница 4: ...roper Measurement 75 4 7 2 Auto Fogging 77 4 7 3 Representative Value 79 4 7 4 REF Confidence Value 79 4 7 5 Standby Mode and Auto Power off 80 4 7 6 Measurement of IOL Eyes 80 4 7 7 If Measured Value...

Страница 5: ...g the Measurement Accuracy 99 7 2 Replacing the Print Roll 99 7 3 Replacing the Fuses 100 7 4 Cleaning the Forehead Rest 101 7 5 Cleaning the Measurement Window 101 7 6 Cleaning Model Eye 101 7 7 Clea...

Страница 6: ...ts around the patient where disturbing measurement it may cause failure of measurement CAUTION Indicates a potentially hazardous situation which if not avoided could result in injury or damage to the...

Страница 7: ...5 no condensation atmospheric pressure 800 1060 hPa This instrument does not meet the temperature requirements of ISO 15004 1 for using Do not use this instrument in conditions where the temperature m...

Страница 8: ...cause afire When replacing the fuse in the station be sure to turn off the power switch and pull out the power cord Be sure to use the designated fuse Use of an undesignated fuse can lead to electric...

Страница 9: ...ith the IEC 60601 1 Anyone who connects additional equipment to the signal input part or signal output part is responsible for making sure that the system complies with the requirements of the IEC 606...

Страница 10: ...dow See 7 5 Cleaning the Measurement Window To clean the measuring window be careful not to make damage or broken of the window screen Internal Battery For clock function lithium battery is equipped a...

Страница 11: ...e printer The battery pack is a consumable item It can be recharged but eventually wears out When the battery pack is no longer usable purchase a new one from your local dealer of the instrument CAUTI...

Страница 12: ...ile radio frequency communication systems because it may have an adverse effect on operation of the device 5 Do not use cables and accessories that are not specified for the device because that may in...

Страница 13: ...are used in such an environment Enclosure port Phenomenon Basic EMC standard or test method IMMUNITY TEST LEVELS Professional healthcare facility environment Electrostatic discharge IEC 61000 4 2 8 kV...

Страница 14: ...ME EQUIPMENT or ME SYSTEM The IMMUNITY TEST LEVELS are applied to the a c power input of the converter 7 Applicable only to ME EQUIPMENT and ME SYSTEMS connected to single phase a c mains 8 E g 10 12...

Страница 15: ...specified frequency range 6 r m s before modulation is applied 7 The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957...

Страница 16: ...1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 2 217Hz 2 0 3 28 2450 2400 2570 Bluetooth WLAN 802 11 b g n FRID 2450 LTE Band 7 Pulse modulation 2 217Hz 2 0 3 28 5240 5500 5785 5100 5800 WLAN 802 1...

Страница 17: ...meter measurement measures the pupil size in the horizontal and vertical directions This instrument consists of a main unit Retinomax Screeen Station Retinomax Station 5 and Printer Retinomax Printer...

Страница 18: ...e installation in such a way that accessible metal parts cannot become live in the event of a failure of the basic insulation Always use the power cord shipped with the instrument It must be connected...

Страница 19: ...nstrument indicates the need for caution This symbol indicates the necessity if referring to the relevant part of the instruction manual before use This symbol indicates that the type of protection ag...

Страница 20: ...Device Directive 93 42 EEC in EU This symbol means the manufacturer This symbol means the manufacture date This symbol means this equipment conforms to China RoHS Management methods for Controlling p...

Страница 21: ...s labels and marks to draw user s attention to important information and precautions If you found some scratch or rack of words please contact to dealer whom you purchased from Main unit Bottom Type B...

Страница 22: ...21 Manufacture date Station Rear of the station...

Страница 23: ...22 Printer Bottom Battery part Manufacture date...

Страница 24: ...23 Battery pack...

Страница 25: ...Forehead rest Center of measuring window Target line for measuring window lateral position Measuring part Right Left eye sensor Target line for measuring window height position Lighting part in CHILD...

Страница 26: ...the power on off Symbol will change depends on the battery pack s capacity Green capacity is over50 Orange capacity is approximately 10 to 50 Blinking with orange capacity under approximately 10 When...

Страница 27: ...ain to release VD Hold down 2 sec Measure with VD 0 Hold down again to release CHILD QUICK Push Switch to CHILD mode and QUICK mode Symbol is blinking CHILD mode lighting QUICK mode ANGLE Refer to rig...

Страница 28: ...ntacts for charging the main unit 4 Accessories mount Contacts for connection to printer 5 Power indicator lamp Charge lamp for main unit Charge lamp for printer Accessories mount Fuse Place the model...

Страница 29: ...ted charging Power indicator lamp OFF Blue Blue Charge lamp for the main unit OFF Orange OFF Charge lamp for the printer OFF Orange OFF 2 4 Printer Indication lamp Paper cover Lever opening closing of...

Страница 30: ...power Blue USB suppling power Light blue Receiving data Pink Error See 8 3 Printer Troubleshooting Pink blinking Data can be output to the PC while USB power is being supplied but it cannot be printe...

Страница 31: ...t in the battery pack with the clicks slide the battery pack to to install 3 To remove the battery pack slide it to reverse direction of 4 To close the grip cover slide it in the opposite direction of...

Страница 32: ...nnect the power cord to AC inlet use power cableenclosed 5 Make sure that the power switch on the station is set to OFF power plug into a power outlet side and plug the 6 On the station set the power...

Страница 33: ...hooting and check the blinking interval If blinking interval is 3 seconds charging stops due to the influence of room temperature Check the room temperature and restart charging if it is between 10 C...

Страница 34: ...hen the charging is completed the charge lamp for the main unit goes out Only use the RT 01XR battery pack which is our designated lithium ion battery Useo fany other battery voids the operationwarran...

Страница 35: ...charge lamp for the printer goes out CAUTION 3 4 3 Charging a Spare Battery Pack Insert the battery pack you want to charge into the printer and charge it See 3 2 2 Printer for how to install the bat...

Страница 36: ...easurement Procedure Display of the opening screen Display of the measurement screen Confirmation of the setting Calibration with model eyes Prepare the forehead rest Positioning the measurement eye B...

Страница 37: ...urement standby screen 4 1 2 Measurement Screen Measurement target Reference and examples of measurement screen Ref measurement number of times Right eye Left eye Stored data sets Storable data sets B...

Страница 38: ...L eyes automatic discrimination Right eye measurement by R L eyes manual setting Left eye measurement by R L eyes manual setting Measurement mode AUTO mode AUTO2 mode Continuous measurement CONT mode...

Страница 39: ...measurement setting OUTPUT Change output setting MEMORY Output and delete stored data CHILD Change CHILD mode setting OTHER Change other setting CLOCK Change time setting POWER OPTION Change power su...

Страница 40: ...S ONE CYL VD 12 0 STEP AUTO OUTPUT OUTPUT SETTING OUTPUT PRINT USB PRINT ID 1 USB FORMAT PC OUTPUT FORMAT REF ALL P S ON MESSAGE OFF EYE OFF CONT FORMAT OUTPUT DATA LAST5 REF ALL P S ON MESSAGE OFF M...

Страница 41: ...N OFF CHART BRIGHTNESS AUTO LCD 80 PARALLEL ON FOCUS ASSIST ON AUTO QUICK OFF CLOCK DATE FORM 120V area MDY 230V area DMY TIME FORM 120V area AM PM 230V area 24H POWER OPTION STAND BY 3MIN SHUT DOWN 3...

Страница 42: ...surement See 4 3 1 Automatic Measurement Mode AUTO AUTO2 to be acknowledged how to measure 4 Compare the values measured with the model eye against the values SPH CYL printed on the model eye Confirm...

Страница 43: ...y be affected If the measured values are out of the above ranges see 4 3 1 Automatic Measurement Mode AUTO AUTO2 and check that your measurement procedure was correct Then retry the measurement Also c...

Страница 44: ...ximately Vertical Align to the target line for measuring window height Lateral Align to the target line for measuring window lateral position Back and forth Position the measuring window approximately...

Страница 45: ...to determine the spectacle lenses that best suit you Please look at the center in a relaxed manner Try to keep your eye as still as possible Example of a tulip fixation target CAUTION In a measuremen...

Страница 46: ...mire ring When the instrument judges that the measurement values are stable the measurement will endautomatically Start Fog REF measurement Measure 5 times at least Auto stop A2 Auto measurement mode2...

Страница 47: ...on the patients eye approximately See 4 2 2 Positioning the Measurement Eye 3 When the is lighting as you move the main unit closer to the patient the automatic right left eye discrimination function...

Страница 48: ...display direction makes it easier to acquire measurement values Move the main body in the direction to be displayed to make it easier to obtain measured values Select OTHER in SETUP and set PARALLEL t...

Страница 49: ...ye measurement If you want to finish the measurement and send the data to printer please press See 4 5 Printing for the data transmission method 4 5 2 Printing Examples for the print example and 5 2 O...

Страница 50: ...7 Storing Data 11 After completed transferring data the case measure the other patient push measurement switch and implement 3 to 9 process A black shadow develops if the pupil has a something that ob...

Страница 51: ...een at measurement CONT 3 If you wanted to end the one eye measurement press the measurement switch Or change to the other eye while measuring 4 When completed measuring both eyes moves on to screen c...

Страница 52: ...ode the symbol will be lighting and CHILD mode will be off CHILD mode The case that measure children sometime they dislike to be measured This instrument contains function that make them pay attention...

Страница 53: ...CHILD mode setting screen CHILD in SETUP See 5 4 CHILD Mode Screen CHILD Perform the measurement in the same process as 4 3 1 Automatic Measurement Mode AUTO AUTO2 and 4 3 2 Continuous Measurement Mo...

Страница 54: ...ful in cases when measurement is difficult because of rapid eye movements due to nystagmus or if the patient is a child or other reason QUICK mod QUICK mode measurement Perform the measurement in the...

Страница 55: ...he right side of the screen Pupil measurement data Pupil size measurement The maximum values of pupil diameter both lateral and vertical directions are measured If the pupil diameter less than 2 0mm p...

Страница 56: ...se the pupil size is less than 2 3 mm 6 When P S is set to ALL measurement is performed 5 times The variation is judged from REF measurement value when the pupil size is 2 3 mm or more and if stable d...

Страница 57: ...e measurement Move the main unit in the indicated directions and measurements will be obtained more easily Measurement screen when alignment is on 1 Enter to SETUP with set ALIGN in OTHER to ON 2 Star...

Страница 58: ...act position 1 Enter to SETUP with and setting with FOCUS ASSIST in OTHER ON Automatically measure when matched focus and alignment DISPLAY Indicates focus No auto measurement OFF The function is off...

Страница 59: ...ing DISPLAY up and down white up and down green up and down green up and down green up and down green up and down green up and down white right and left blue right and left green right and left orange...

Страница 60: ...e horizontal indication display is displayed in 2 increments Measurement screen when horizontal correction is ON 3 When you wanted to release the auto correction hold down for 2 seconds again 4 When t...

Страница 61: ...you need to perform measurement from 45 90 or 135 rotated position as for bedside measurement astigmatic axis AX must be set appropriately The rotation correction function corrects the astigmatic axis...

Страница 62: ...rection of the arrow with patient shead 3 To be off rotation correction function hold down for 2 seconds again Rotation corrected angle select screen 4 Push the arrow changes form blinking to lighting...

Страница 63: ...Each corrected angle for right left eye measurement is displayed on measurement completed screen Rotation corrected angle Measurement completed screen with rotation correction function Printing examp...

Страница 64: ...ed after 6 seconds shift to QUICK mode automatically 3 When switching to QUICK mode AQ is displayed on the left side of the screen Measurement screen when switching to QUICK mode by AUTO QUICK mode Th...

Страница 65: ...64 5 When shifting to QUICK mode by AUTO QUICK mode AQ is printed on the measured data Printing example...

Страница 66: ...ing and mire ring and background illumination are light out only the central cornea reflected light and pupil can be seen Any foreign objects such as opacity in the crystal lens that obstructthe measu...

Страница 67: ...memory function is on the transferred data to the printer automatically stored to the main unit 1 Hold down for 2 seconds on measurement standby screen lights and memory function is on 2 After measur...

Страница 68: ...measurement standby screen again 2 Quick Saving This function can manually save the measured data when the memory function is off 1 Implement measurement with the memory function off 2 Hold down for 2...

Страница 69: ...68 4 Printing the data stored by quick saving QS is printed on the printing data Printing example one eye after measured...

Страница 70: ...RMAT with SETUP to ALL an error will occur if data is transmitted while the main unit and the printer are separated by 50 cm or more Be sure to transmit the distance between the main unit and the prin...

Страница 71: ...nit There is an angle of more than 45 degrees between the front side of the main unit and the data receiving window of the printer In such cases move near the printer and press once again to send data...

Страница 72: ...values Right eye REF representative value Right eye Spherical equivalent power Right eye Pupil size representative value in representative value REF confidence value Left eye REF pupil diameter values...

Страница 73: ...4 5 3 Cutting off the Printer Paper Auto cutter is equipped on the printer and the paper is automatically cut Tear out the paper by pulling it toward you since the central of the paper is still tied...

Страница 74: ...T SETTING see 5 2 1 Output Setting OUTPUTSETTING If PRINT ID is set to SELECT the screen switches to the PRINT ID SELECT screen each time when data is transmitted Please select the PRINT ID set for th...

Страница 75: ...down for 2 seconds with sub menu in SETUP Hold down for 2 seconds and release in standby mode Power off with the setting of Auto Power Off Place the main unit to the station with power on When the ba...

Страница 76: ...ely in front of the patient s face as shown below The discrimination may fail if measurement is taken at an angle upward or sideways If you follow the above main unit but the right left eye discrimina...

Страница 77: ...e lights off When you wanted to measure right eye press andindicatedrighteyeand symbol turns on When you wanted to measure left eye press and indicates left eye and symbol turns on If you want to rest...

Страница 78: ...Fogging Fixation target appears blurred to the patient 4 Obtain data for 5 times When data is obtained a beep sound is made Fixation chart remains in fogging state and the appears blurred to the pati...

Страница 79: ...ained 7 If the measurement values are fluctuating repeat movement of fixation chart and 1 data obtainment for 3 times until the measured value is stabilized The main unit stores up to 8 times of data...

Страница 80: ...lease use the following as a guide for confidence value 8 or higher Low variation 7 or lower High variation Care is needed in interpreting the measurement results Pleasemeasure again with focus assist...

Страница 81: ...ent errors If it is difficult to obtain measurements using the Quick Mode may make the measurement easier If the auto quick function is set beforehand it automatically changes to Quick Mode See 4 4 5...

Страница 82: ...g an eye with a contact lens Measurements can usually be obtained However if the contact lens is not properly fitted correct values may not be obtained Dirt or scratch on the contact lens may hinder m...

Страница 83: ...7 POWER OPTION Screen POWER OPTION PATIENT No To 5 8 Patient Number Setup PATIENT No MESSAGE To 5 9 MESSAGE Screen MESSAGE INITIAL To 5 10 INITIALIZING Screen INITIAL VERSION To 5 11 VERSION Confirma...

Страница 84: ...surement ALL Measuring pupil diameter each time with REF measurement OFF Not measuring the pupil diameter CYL MIX Without CYL measurement value conversion Converts CYL value to minus Converts CYL valu...

Страница 85: ...en to configure external output setting EXIT Back to the SETUP screen OUTPUT SETTING To 5 2 1 Output Setting OUTPUT SETTING OUTPUT FORMAT To 5 2 2 Output Format OUTPUT FORMAT CONT FORMAT To 5 2 3 Outp...

Страница 86: ...ction Setting EXIT Back to the OUTPUT screen OUTPUT PRINT ONLY Implement print only PRINT USB Implement printing and transfers data to PC via USB USB ONLY Implement only transferring the data to PC vi...

Страница 87: ...OUTPUT screen REF ALL Printout all of measurement data maximum is 8 and representative value REP Printout only representative value OFF Not printout REF value P S ON Printout pupil diameter OFF Not pr...

Страница 88: ...LAST5 The last 5 data of REF measurement values are output ALL Output all REF measurement values Up to 100 data can be output REF ALL Output REF measurement value and representative value set with OU...

Страница 89: ...sfer 30 latest person s data at maximum PRINT Printout the stored data The number of transfer depends on which you set at OUTPUT NUMBER USB Output stored data by USB The number of transfer depends on...

Страница 90: ...when you press 3 The data transferred when you press again 4 When you press the masurement switch it goes back to MEMORY NUMBER SELECT 2 After selected DELETE screen indicates confirmation If selecte...

Страница 91: ...elody is made at CHILD mode OFF Melody isn t made at CHILD mode LED ON LED at peripheral measurement window lights at CHILD mode OFF LED at peripheral measurement window doesn t light CHART COLOR ON V...

Страница 92: ...IGHTNESS HIGH Brighten the fixation target LOW Darken the fixation target AUTO Brightness of target automatically changes If pupil diameter is over 3mm HIGH Within 3mm LOW LCD 5 to 100 Adjust brightne...

Страница 93: ...the SETUP screen DATE FORM YMD year month day Changes the date print format MDY month day year D DMY day month year TIME FORM AM PM 12 hour system D Changes the time print format 24 hour system Selec...

Страница 94: ...cally Item Selection Setting EXIT Back to the SETUP screen STAND BY 1 to 9 MIN NEVER Set time to switch into standby mode Press button to add 1minute Press button to minus 1 minute SHUT DOWN 1 to 9 MI...

Страница 95: ...9 Sets the 1st digit key adds one key subtracts one 0010 0 to 9 Set the 10th digit key adds one key subtracts one 0100 0 to 9 Sets 100th digit key adds one key subtracts one 1000 0 to 9 Sets 1000th di...

Страница 96: ...can be as long as 24 characters by 2 lines or 48 characters in total How to enter and set a message Button Operation Move the cursor relative position Measurement switch Enter the character at the cur...

Страница 97: ...screen When you selected GO confirmation message on the screen Press INITIAL EXEC initialize all data in the SETUP and clears memory After completed FINISHING INITIALIZATION on the screen in 3 sec and...

Страница 98: ...VERSION VERSION can check the software version Select PRINT USB the main unit toward to the printer and press to print the software version of the main unit and the software version of the printer Se...

Страница 99: ...Remote Vision RV II the case of environment they positioned In order to wirelessly transmit to RV II it is necessary to upgrade RV II When upgrading RV II please contact the place of purchase 6 2 Conn...

Страница 100: ...ease lever 2 Remove the remaining print roll 3 Remove the shipping tape from the new print roll 4 Put a crease slightly past the peel mark 5 Gently cut the paper along the crease Cutting out the paper...

Страница 101: ...20 mm 250V T1 6AH 021501 6XP CAUTION 7 3 Replacing the Fuses CAUTION If the power lamp does not light up when the power switch of the station is set to ON I position the fuses may have blown Remove th...

Страница 102: ...ed with the model eye its lens surface may have been contaminated by dust or fingerprints In this case gently wipe the lens surface using a soft clean cotton cloth such as gauze moistened slightly wit...

Страница 103: ...s time lug fuses 021501 6XP Rating 250V 1 6A Size 5 20mm Battery pack RT 10XR Lithium ion battery 7 9 Lifetime of the Instrument Expiration date of this instrument is 8 years except consumable parts a...

Страница 104: ...y switches off Charge the battery Are the batteries charging contacts 4 places dirty See 2 1 Main Unit Turn the power off and clean the pins both on the main unit and the station sides Is the main uni...

Страница 105: ...nter is possibly not set to printout Make sure the setting See 5 2 OUTPUT Screen OUTPUT Error message If the following message is displayed and repeated again even after the equipment is turned off an...

Страница 106: ...arging Lamp The case that charging lamp flashes while it charging defective of charging is considered The procedure to make sure following The charge lamp flashes at 0 5 sec intervals Charge lamp ligh...

Страница 107: ...ndow and transfer again The printer does not receive measurement data sent from the main unit Are there any obstacles between the main unit and printer Remove the obstacles or move to a location free...

Страница 108: ...s a lithium primary battery If the battery exhausted the date time varies If the display as shown in the figure following appears please contact to the dealer where you purchased the equipment Display...

Страница 109: ...tex distance 12 13 5 13 75 15 16 0 Pupil size measurement Measurement range 2 0mm to 12 0mm Measurement step 0 1mm Other Functions Automatic right left eye detector Alignment direction indication Focu...

Страница 110: ...month day time minute Monitor 3 5 inch color LCD QVGA External output Communication to Retinomax Printer 5 via Infrared max 6m Fixation chart Tulip Bear Firework factory option Fixation chart lighting...

Страница 111: ...Powered from the station when docked with the station Battery drive when separated from the station and abattery pack is installed USB supply from the computer when separated from the station and with...

Страница 112: ......

Отзывы: