4-1-4 Measuring the Model Eye
1.
Open the accessory case, and, place model eye.
2.
Make sure setting with AUTO mode and VD12.0mm.
3.
Focus on model eye and start measurement.
*See “4-3-1 Automatic Measurement Mode (AUTO, AUTO2)” to be acknowledged
how to measure.
4.
Compare the values measured with the model eye against the values (SPH, CYL)
printed on the model eye. Confirm that the differences are within the following ranges:
SPH
:
±0.25 D
CYL
:
±0.25D
Exposure to light may lead to loss of measurement accuracy or failure of the
measurement.