8
•
Do not put any objects around power switch of the station or the printer to not
disturbing emergency operation.
•
Do not lean on the units.
•
Please clean according to the specified method on the manual. If not follow, it
may cause deterioration or damage to the unit. (See “7-4 Cleaning the
Forehead Rest” to “7-7 Cleaning Appearance”)
•
Clean the forehead rest for every patient, if not, it may cause cross infection.
•
For repair or maintenance, wipe the outer cover with cloth with alcohol for
sterilization.
•
When the device is return for maintenance or repair, clean the surface of the
device (especially, forehead rest) with a cloth dampened with rubbing alcohol
for disinfections.
•
Accessory equipment connected to the analog and digital interfaces must be
certified according to the appropriate national standards. Furthermore, all
configurations shall comply with the IEC 60601-1. Anyone who connects
additional equipment to the signal input part or signal output part is
responsible for making sure that the system complies with the requirements of
the IEC 60601-1.
•
The operator must not touch the external instrument connected with this
equipment and the patient simultaneously. Electrical shock can result.
•
If in doubt, consult the technical service department or your local
representative.
CAUTION
6. Transportation / Storage
•
The ambient conditions of transportation: Temperature -20°C to 60°C, relative
humidity 10% to 95% (no condensation), atmospheric pressure 500-1060
hPa.
•
The ambient conditions of storage in a packaged condition: Temperature
-10°C to 45°C, relative humidity 10% to 85%(no condensation), atmospheric
pressure 700-1060 hPa.
•
This instrument does not meet the temperature requirements of ISO 15004-1
for storage. Do not store this instrument in conditions where the temperature
may rise above 45°C or fall below -10°C.
•
Select a storage area exposed to limited amounts of dust, free of vibration and
shock.
Содержание Retinomax K+Screeen
Страница 1: ...Hand Held Auto Refractometer Instructions November 2018 D054 E...
Страница 22: ...21 Manufacture date Station Rear of the station...
Страница 23: ...22 Printer Bottom Battery part Manufacture date...
Страница 24: ...23 Battery pack...
Страница 65: ...64 5 When shifting to QUICK mode by AUTO QUICK mode AQ is printed on the measured data Printing example...
Страница 112: ......