5-1
Measurement Setup [MEASURE]
Use this screen to configure setting that work during measurement.
Item
Selection
Setting
EXIT
-
Back to the [SETUP] screen.
TYPE
AUTO
Switches to AUTO mode.
AUTO2
Switches to AUTO2 mode.
CONT
Continuous measurement mode.
P-S
ONE
Measuring pupil diameter once with REF measurement.
ALL
Measuring pupil diameter each time with REF measurement.
OFF
Not measuring the pupil diameter.
CYL
MIX
Without CYL measurement value conversion.
-
Converts CYL value to minus.
+
Converts CYL value to plus.
VD
12.0
Sets corneal vertex distance (VD) in mm.
13.5
13.75
15.0
16.0
STEP
AUTO
Display the measurement value in 0.12D steps if measurement
value within
±
3D. The other displays in 0.25D steps.
0.25
Display the measurement value in 0.25D steps.
0.12
Display the measurement value in 0.12D steps.