70
1.
Set the print roll correctly. (See “7-2 Replacing the Print Roll”)
2.
Make sure the printer power on.
3.
After the end of the measurement, press
toward the printer’s data receiving
window, the measurement data successfully transferred.
4.
The printer discharges print paper showing the measured value after transfer.
The print content is output according to the setting of “5-2 Output setting screen
[OUTPUT]”.
・
The printer may fail to operate in the following situations:
◆
The light from the main unit does not reach the printer when, for example, a
person passes between the main unit and printer.
◆
The printer is used at a place more than 6 m away from the main unit.
◆
There is an angle of more than 45 degrees between the front side of the
main unit and the data receiving window of the printer.
・
In such cases, move near the printer and press
once again to send
data. Printout can be made as many times as necessary before the next
measurement starts.
・
When measuring the next patient, be sure to make a printout before starting
the measurement.
・
Retinomax Screeen series cannot be used with a printer provided for use with
Retinomax, Retinomax 2, or Retinomax 3.
・
If print paper runs out during printing, the indicator light blinks in pink.
Replace the printing paper according to “7-2 Replacing the Print Roll”. After
replacing printing paper, automatically reprint the data. After the indicator
blinks in pink, data is deleted after 5 minutes without replacing the print paper.
Please push
again and transmit measurement data.
Содержание Retinomax K+Screeen
Страница 1: ...Hand Held Auto Refractometer Instructions November 2018 D054 E...
Страница 22: ...21 Manufacture date Station Rear of the station...
Страница 23: ...22 Printer Bottom Battery part Manufacture date...
Страница 24: ...23 Battery pack...
Страница 65: ...64 5 When shifting to QUICK mode by AUTO QUICK mode AQ is printed on the measured data Printing example...
Страница 112: ......