Cleaning the Forehead Rest
Cleaning should be done before the measurement. Clean the forehead rest with soft cloth,
etc. soaked with rubbing alcohol.
7-5
Cleaning the Measurement Window
The measuring window has dust-proof glass. If any dust on the glass is visible from the
patient side, blow off the dust three to four times using the blower provided. If the dust
cannot be removed, wipe off using a lens cleaning solution or absolute alcohol (commercial
product).
7-6
Cleaning Model Eye
When correct measured values cannot be obtained with the model eye, its lens surface may
have been contaminated by dust or fingerprints. In this case, gently wipe the lens surface
using a soft, clean cotton cloth (such as gauze) moistened slightly with lens cleaning
solution or ethyl alcohol for disinfections, being careful not to scratch the surface.
Do not use a handkerchief or absorbent cotton.
7-7
Cleaning Appearance
If the appearance stained, wipe off with soft cloth, or the case that stubborn stains on the
appearance, wipe off lightly with a gauze moistened slightly with diluted neutral detergent.
As protection, utilizing dustcover on the instrument after measurement is recommended.
against the model eye as this may cause damage.
discoloration, or peeling of printed characters.
Содержание Retinomax K+Screeen
Страница 1: ...Hand Held Auto Refractometer Instructions November 2018 D054 E...
Страница 22: ...21 Manufacture date Station Rear of the station...
Страница 23: ...22 Printer Bottom Battery part Manufacture date...
Страница 24: ...23 Battery pack...
Страница 65: ...64 5 When shifting to QUICK mode by AUTO QUICK mode AQ is printed on the measured data Printing example...
Страница 112: ......