105
8-2
Station
Phenomenon
Check point
Cause and action
The charge lamp
flashes during
charging.
Long
(Flashes lights 3 sec, off 0.5
sec)
Preservation function works
when the battery thermal has
increased or decreased.
Check the temperature in the
room (10°C to 35 °C). When
the temperature decreased
(10°C to 35 °C), restarts
charging.
Short
(Flashes each of 0.5 sec)
Re-connect the main unit or
printer regarding flow
following.
If you considered it were
defect, contact to the dealer.
●Check Flashing of the Charging Lamp
The case that charging lamp flashes while it charging, defective of charging is considered.
The procedure to make sure following:
The charge lamp flashes at 0.5 sec
intervals.
Charge lamp lights continuously.
The charge lamp flashes at 0.5 sec intervals.
Remove the main unit or the printer from the station.
The charge lamp lights continuously.
The charge lamp flashes at 0.5 sec
intervals.
It is likely that the contact condition
was improper during the initial
charging.
Continue charging.
Replace with another normal battery
pack and set up for charging.
There is likely a problem with the
station. Contact your dealer.
There was likely a problem with the
old battery pack.
Reset for charging.
Содержание Retinomax K+Screeen
Страница 1: ...Hand Held Auto Refractometer Instructions November 2018 D054 E...
Страница 22: ...21 Manufacture date Station Rear of the station...
Страница 23: ...22 Printer Bottom Battery part Manufacture date...
Страница 24: ...23 Battery pack...
Страница 65: ...64 5 When shifting to QUICK mode by AUTO QUICK mode AQ is printed on the measured data Printing example...
Страница 112: ......