Fixed calibration
: Means to execute tool setting operation at a certain fixed position on the
machine table.
During machining, the cutter tool wear or fracture that needs replacement. However, the blade
length and tool clamping position will change after tool changing. The system can execute “fixed
calibration” to re-confirm the tool offset value.
Instructions
:
1)
Set microset position
Get into “MACHINESETUP--C.A.D Position” to set microset position, including
“Inplace”&“Inposition”.
In position(positioning): Press“
”, screen displays “Press OK key to set point position by
manual mode, it’s very simple but not accurate, press cancel to set by number”. Recommend to
choose manual mode.
In place: Move X,Y axis to the place where is over the microset, press “
”.
2)
First time tool setting / Tool setting after tool changing
Manual move Z axis to the surface of workpiece, and set workpiece origin of X,Y,Z axis.
Press “
”execute the first time tool setting. The system will record value of offset
automatically.
Start the processing after first time tool setting.
After the tool changed or tool bit broken, press “
”+“
” to start second time tool
setting, system will automatically restore to current Z axis workpiece coordinate origin.
Second time tool setting completed, start processing.
Note: After the tool changed or tool bit broken, the system can automatically calculate Z offset
and save it when executes the tool setting command, after tool setting user can start
processing directly without press “Z
→
0”.
PS5.3 Mobile calibration
Mobile calibration is suitable for materials such as wood working, acrylic, plastic, aluminum, iron,
steel etc.
Generally, the tool touch-off plate can be used to do the mobile calibration, as it is economical,
simple and convenient.
Tool touch-off plate wiring:
1. The white wire connects to X4(TS) input port on the wiring board.
2. The black wire connects to GND on the input terminal of the wiring board.