Operation
© Rice Lake Weighing Systems
●
All Rights Reserved
51
Reading Weight, The Setpoint And The Peak (If Present) From The Pc:
The PC transmits:
$aajckckCR
To read the points, set the
equal to 50000. $01p71CR
Possible instrument responses:
• Correct reception:
&aaxxxxxxj\ckckCR;
in which
xxxxxx
= 6 value characters of the required weight
• Incorrect reception:
&&aa?\ckckCR
• If the peak is not configured:
&aa#CR
In the event of negative weight, the first character on the left acquires the value « - » (minus sign - ASCII 45).
In the event the weight value is under -99999, the minus sign (‘-‘) is sent alternated with the most significant figure.
Error messages
In the event of an instrument alarm for exceeding 110% of the full scale or nine divisions above the value of the parameter
, the instrument sends the string
&aassO-Lst\ckck;
in which: s = 1 separator character (32 ASCII – space-).
In the event of a faulty connection of the load cells or other alarm, the instrument sends
&aassO-Fst\ckck;
in which:
s
= 1
separator character (32 ASCII – space-).
See
for more information on alarms.
Semi-Automatic Zero (Weight Zero-Setting For Small Variations)
The zero-setting will not be maintained after an instrument power-off.
The PC transmits:
$aaZEROckckCR
$01ZERO03CR
Possible instrument responses:
• Correct reception:
&&aa!\ckckCR
• Incorrect reception:
&&aa?\ckckCR
• The current weight is over the maximum value resettable:
&aa#CR
Switching From Gross Weight To Net Weight
The PC transmits:
$aaNETckckCR
$01NET5ECR
Possible instrument responses:
• Correct reception:
&&aa!\ckckCR
• Incorrect reception:
&&aa?\ckckCR
Switching From Net Weight To Gross Weight
The PC transmits:
$aaGROSSckckCR
$01GROSS5BCR
Possible instrument responses:
• Correct reception:
&&aa!\ckckCR
• Incorrect reception:
&&aa?\ckckCR
Data
Definitions
j
a to read setpoint 1
$01a60CR
j
b to read setpoint 2
$01b63CR
j
c to read setpoint 3
$01c62CR
j
t to read gross weight
$01t75CR
j
n to read net weight
$01n6FCR
j
p to read the gross weight peak if the ASCII parameter is set as NOdU60;
if the ASCII parameter is set on NOd td the gross weight will be read.
Table 3-42. Weight, Setpoint, Peak Definitions (ASCII)
Note
IMPORTANT