SCT-20 Series
28
Visit our website
3.0
Operation
3.1 Semi-automatic Tare (Net/Gross)
The semi-automatic tare value is lost upon instrument power-off.
The semi-automatic tare operation is not available if the gross weight is zero.
1. To capture tare and weigh in net mode (Semi-Automatic Tare), close the Net/Gross input or press for three
seconds. The instrument displays the net weight (zero) and the
NET
LED lights up.
2. To display the gross weight again, keep the Net/Gross input closed or press for three seconds.
This operation may be repeated by the operator to allow the loading of several products.
Press and hold to display the gross weight temporarily. When is released, the net weight will be displayed
again.
3.2 Preset Tare (Subtractive Tare Device)
It is possible to manually set a preset tare value to be subtracted from the display value provided that the
is less than
max capacity.
1. Press and hold
and .
is displayed
Press
.
2. Press or until desired value is displayed. Press
.
3. Press to exit
. After setting the tare value, the weight display shows the net weight (subtracting the preset
tare value) and the
NET
LED lights up to show that a tare has been entered.
Press and hold for three seconds to display the gross weight temporarily. When is released, the net weight
will be displayed again.
To delete a preset tare and return to the gross weight display:
Press and hold for three seconds or keep the Net/Gross input closed for three seconds. The preset tare value is set to zero.
The
NET
LED is turned off when the gross weight is displayed once again.
If a semi-automatic tare is entered, it is not possible to access the preset tare function.
If a preset tare is entered, it is still possible to access the semiautomatic tare function. The two different types of tare
are added.
All the semi-automatic tare and preset tare values will be lost when the instrument is turned off.
3.3 Semi-automatic Zero (Weight Zero-setting For Small Variations)
By closing the
Semi-Automatic Zero
input, the weight is set to zero. The zero setting will be lost when the instrument is turned
off.
This function can only be used if the weight is lower than the 0 set value, otherwise the
t
-----
alarm appears and the weight is
not set to zero. See
3.4 Peak
By keeping the input closed, the maximum weight value reached remains displayed. Upon opening the input, the current weight
is displayed.
To view a sudden variation peak, set the FILTER ON THE WEIGHT to 0.
Note
Note
Note
Note
Note