3.15 Interface to Remote Display
Remote Display (Laser Light)
1. Press and hold
, then press .
is displayed.
2. Press or until
is displayed.
3. Press
.
is displayed.
4. Press
.
is displayed.
5. Press or until
is displayed. Press
.
6. Press twice to exit set-up menu.
Laser Light
E CHAR = CR
LWPOS = 7
LENGTH = 19
Laser Light 1550
S CHAR = SoH
LWPOS = 000008
3.16 Communication Examples
The numerical data below are expressed in hexadecimal notation with prefix h.
Example 1: Command for multiple writing of registers (hexadecimal command 16, h10)
Assuming that we wish to write the value 0 to the register 40017 and the value 2000 to the register 40018, the string to generate
must be:
h01 h10 h00 h10 h00 h02 h04 h00 h00 h07 hD0 hF1 h0F
The instrument will respond with the string:
h01 h10 h00 h10 h00 h02 h40 h0D
Query Field Name
Hex
Response Field Name
Hex
Instrument Address
h01
Instrument Address
h01
Function
h10
Function
h10
Address of the first register H
h00
Address of the first register H
h00
Address of the first register L
h10
Address of the first register L
h10
Number of registers to send H
h00
Number of registers H
h00
Number of registers to send L
h02
Number of registers L
h02
Byte Count
h04
CRC16 H
h40
Datum 1 H
h00
CRC16 L
h0D
Table 3-52. Communication Example 1
Laser Light
SCT
Connector
Pin
Connector
J8 (RS-232)
3
RS-485 RX-
5
RS-485 RX+
Table 3-51. Laser Light Setup