2.1.2 Theoretical Calibration
This function allows load cell rated values to be set. To perform the theoretical calibration set
,
and
in
sequence. See
Figure 2-2 on page 10
.
1. Press and hold
, then press .
is displayed.
2. Press
.
is displayed. Press
.
3. Press or until total load cell capacity (system full scale) is displayed. Press
.
4. Press or until
is displayed. Press
.
5. Press or until desired load cell mV/V is displayed. Press
.
6. Press or until
is displayed. Press
.
7. Press or until desired display division size is displayed. Press
.
8. Press twice to exit setup menu.
By modifying the theoretical full scale, the sensitivity or divisions, the Weight (Span) Calibration is canceled and only
the theoretical calibration is considered valid.
If the theoretical full scale and the recalculated full scale in Weight (Span) calibration (
) are
equal, this means that the calibration currently in use is theoretical calibration; if they are different, the calibration in
use is the Weight (Span) Calibration based on test weights.
By modifying the theoretical full scale, the sensitivity or divisions and all the system’s parameters containing a
weight value will be set to default values (setpoints, hysteresis, etc.).
2.1.3 Maximum Capacity (
)
Maximum capacity is the highest weight of live load/product that can be displayed. When the weight exceeds this value by 9
divisions, the following is displayed:
------
. To disable this function, set to 0.
1. Press and hold
, then press .
is displayed.
2. Press
,
is displayed.
3. Press or until
is displayed. Press
.
4. Press or until desired capacity is displayed. Press
.
5. Press twice to exit setup menu.
2.1.4 Zero Setting
Use Zero Setting to zero the scale after use or to compensate for variations due to environmental factors.
Perform this procedure after having set the theoretical calibration. See
.
This menu may also be accessed directly from the weight display. Press and hold for three seconds.
1. Press and hold
, then press .
is displayed.
2. Press
.
is displayed.
3. Press or until
is displayed. Press
. The weight value to be set to zero is displayed and all of the
LEDs are flashing.
4. Press
, the weight is set to zero and the value is stored to the permanent memory.
5. Press twice to exit set-up menu.
Press to display the accumulated deadload that has been zeroed off by the instrument, displaying the sum of all
of the previous zero settings.
Note