Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
15
All other settings are expressed in weight value. To convert them to the new unit of measurement, perform a
Theoretical Calibration For Other Units Of Measure on page 15 or a Weight (Span) Calibration For Other Units Of
Measure on page 15 to change the system calibration. The parameter
must remain set to 1.0000.
Theoretical Calibration For Other Units Of Measure
Set in the parameter the
value divided by the conversion coefficient from kg to the new unit of measure.
Example: The 4 load cells of 1000 kg are placed under a scale for olive oil, which has a specific gravity of 0.916 kg / l. Setting
the
= (4x1000) / 0916 = 4367, the system works in liters of olive oil. See
Weight (Span) Calibration For Other Units Of Measure
Load a known quantity of product litres on the scale (use as high a percentage of the maximum quantity to be weighed as
possible) and enter in the parameter
, the product loaded value in litres. See
2.2 Filter On The Weight
The filtering selection is used to eliminate environment noise, and is typically a compromise between responsiveness and
stability. The lower the number, the more responsive the display will be to weight changes. The filter is used to stabilize a weight
as long as the variations are smaller than the corresponding Response Time. The filter setting is dependent on the type of
application and the required update rate (see
.
Setting the parameter allows a stable weight display to be obtained. To increase the effect (weight more stable), increase value.
1. Press and hold
, then press .
is displayed.
2. Press or until
is displayed.
3. Press
. The currently programmed filter value is displayed.
4. Press or until desired filter value is displayed.
5. Press
to accept. The current weight will then display with all LEDs scrolling and the displayed stability can be
verified.
• If the stability is not satisfactory, press twice. This returns the indicator to the
option and the filter may
be modified again until an optimum result is achieved.
• If the stability is satisfactory, press
. Do not proceed until
is displayed.
• Press once to set the anti-peak filter.
The anti-peak filter can only be set after the instrument is connected to a load cell. When the weight is stable, the
anti-peak filter removes sudden weight fluctuations determined by the configured time (see
), up to a
maximum duration of one second.
6. Press or to display
or
to set this filter on or off
7. Press
to accept.
is displayed.
8. Press to exit setup menu.
Filter Value
Response Time [ms]
Display and Serial Port Refresh Frequency [Hz]
0
12
300
1
150
100
2
260
50
3
425
25
4*
850
12.5
5
1700
12.5
6
2500
12.5
7
4000
10
8
6000
10
9
7000
5
Anti-peak
6
600
* Indicates default value.
Table 2-5. Filter Values
IMPORTANT
Note