Operating the machine
23
3474s/- NORMAL
o
Opening the door interrupts the size reduction process and the main menu is
shown again.
CONT. NORMAL
READY
NOTE
If the plates get stuck (hard sample), the device switches automatically into reverse
gear to start the plates moving again with maximum momentum. After 5
unsuccessful automatic attempts, the process is stopped.
5.7.2
Faults during Size Reduction
Fault
Cause and possible remedy
Sample material escapes
Check contents in bag (samples that are too hard,
e.g. shells, can damage the bag). Examine the bag
quality (e.g. for weak seams). If the fault occurs
repeatedly, please ask customer service.
Message
“Size reduction failed”
Take the bag out and check if the sample is not too
large.
If the reason for the message is not apparent, please
read through the “Fault diagnosis” section.
Note: When technical faults occur, please refer to the error messages section.
Pos : 6. 21 /00005 Ü berschrif ten/ 1. 1 Ü bersc hrift en/ 1. 1 Übersc hrift en BD A/ 11 Durc hlauf dauer @ 2\ mod_1263289874150_9.doc @ 18528 @ 2
5.8 Process Run Duration
Pos : 6. 22 /00010 Bedi enungsanleit ung en/00140 Kapit elsamml ung Paddle Bl ender/0003 PB Kapit el Bedienung/ 0309 PB M odul D urchl aufdauer @ 2\ mod_1263290871222_9.doc @ 18556 @
The size reducer can run either in continuous operation or for a time which can be
adjusted from 10 - 180 sec.
60s NORMAL
READY
Starting with the main screen, set a process run duration of
XXXs
(flashes) with the
◄ or ►
key.
When put into operation for the first time, the device’s default setting is a
process run duration of 60 sec.
60s NORMAL
READY
Use the
▲
or
▼
keys to scroll through the list of the size reduction times,
ending with
continuous operation
.
Tests have shown that a process run duration of 30 - 60 sec. is sufficient
for the size reduction of most samples.
Pos : 6. 23 /00005 Ü berschrif ten/ 1. 1 Ü bersc hrift en/ 1. 1 Übersc hrift en BD A/ 11 Motor drehzahl @ 1\ mod_1241508280705_9.doc @ 8763 @ 2
5.9 Motor rotation speed
Pos : 6. 24 /00010 Bedi enungsanleit ung en/00140 Kapit elsamml ung Paddle Bl ender/0003 PB Kapit el Bedienung/ 0310 PB M odul M ot ordr ehz ahl @ 2\ mod_1263291329108_9. doc @ 18563 @ 3
60s NORMAL
READY
Starting with the main screen, use the
◄ or ► keys to set a
motor
speed (flashes).
60s NORMAL
READY
Use the
▼
or
▲
keys to scroll through the list of speeds (normal, slow, fast).
When put into operation for the first time, the default setting is “normal”
speed.
5.9.1
Plate movements per minute depending on speed
Speed
Movements per min.
Movements per min.