Cleaning and service
27
• Lubricate the fulcrum pins with Vaseline oil once a month.
Fig. 13: Lubricating the Fulcrum Pins
Pos : 7. 9 / 00005 Übersc hrift en/1. 1 Ü berschrif ten/ 1. 1 Ü bers chrift en BDA/ 11 Aus tausc h der Kammer dichtung @ 2 \ mod_1262961360954_9. doc @ 18481 @ 2
6.3 Replacing the chamber sealing
Pos : 7. 10 /00010 Bedi enungsanleit ung en/00140 Kapit elsamml ung Paddle Bl ender/0004 PB R einigung/0403 PB M odul Aus tausc h der Dichtung @ 2\ mod_1262961411105_9. doc @ 18488 @
Fig. 14: Replacing the sealing
NOTE
The sealing on the size reduction chamber must be replaced at least once a year
and at shorter intervals when damaged.
It is recommendable to leave the door open a bit when the blender is not in use so
that the sealing does not stick together or get warped.
• Before inserting the sealing, clean the sealing groove with 70% alcohol.
• Put the two ends of the sealing into the upper parts of the groove (
1
) (
2
).
• Put the middle part of the sealing into the lower part of the groove (
3
).
• Push the remaining sealing with even tautness into the groove (
4
) (
5
).
Pos : 8. 1 / 00010 Bedienungsanlei tungen/------- Seit enumbruc h ----------- @ 0\ mod_1222344373758_0. doc @ 2386 @
1
2
3
4
5