RM-111 Operations Manual
Application 13
Application 13
9. Make proper connections to the RM-111, see fi gure 19.0. Connect Pulse IN(s) to port(s) A, B,
and/or D and Pulse OUT to port D.
10. Set up the test board to test the solid state meter. Include all variables such as: form number,
volts, current, and number of pulses per revolution. If necessary, specify that a through-hole test
will be conducted.
11. Start the test on the test board. The RM-111 starts receiving pulses. After the specifi ed number
of pulses has been counted the RM-111 outputs one pulse. This continues until the RM-111’s
CANCEL key is pressed.
12. The RM-111’s display shows the pulses IN per the pulses OUT for as long as the test is ran.
For this application there are no results and therefore no save option.
84