RM-111 Operations Manual
KH - Enter the DUT Kh value.
PriKh – Enter the primary reference standard(s) Kh value.
Test by* – Select either Revs or Energy.
#Standards – Enter the number of reference standards that will be used.
Mode – If more than one reference standard is used, the user must choose between single-phase or
polyphase modes of operation.
Ref Type – This defi nes the reference standard output pull-up value. Select Radian (150 ohm).
Other is 1000 ohm.
#Inputs – Enter the number of current inputs used for the reference standard(s).
SER#1 – Enter a serial number for standard 1, if applicable.
SER#2 – Enter a serial number for standard 2, if applicable.
SER#3 – Enter a serial number for standard 3, if applicable.
Element – Enter the number of meter elements.
FL Revs – Enter the desired number of revolutions for the full load test (FL).
PF Revs – Enter the desired number of revolutions for the power factor test (PF).
LL Revs – Enter the desired number of revolutions for the light load test (LL).
FL WAVG – Enter the desired weight of the measured FL value in the weighted average
calculation.
PF WAVG – Enter the desired weight of the measured PF value in the weighted average
calculation.
LL WAVG – Enter the desired weight of the measured LL value in the weighted average
calculation.
Demand – Enter the desired number of minutes for the demand test. To omit the demand test,
enter zero.
Application 7
Application 7
58
Содержание RM-111
Страница 1: ...Operations Manual Radian Research Inc RM 111 Automated Comparator...
Страница 3: ......
Страница 8: ...RM 111 Operations Manual Product Introduction Product Introduction 7...
Страница 30: ...RM 111 Operations Manual Figure 4 0 Testing a Radian Standard Counting Pulses Application 1 Application 1 29...
Страница 39: ...RM 111 Operations Manual Figure 6 0 Testing a Radian Standard Display Gating Application 3 Application 3 38...