RM-111 Operations Manual
Name - Enter a test setup name on the fi rst line by typing over the current text (16 characters
maximum).
PriKh – Enter the primary reference standard(s) Kh value.
Kh – Enter the D.U.T. Kh value.
#Pulse – Enter the number of pulses for the reference standard(s) to output before stopping the
test.
Time* – not applicable for this test
Test by* – Select PULSE
Type – Select Cnt Pulse for the gating method
#Standards – Enter the number of reference standards that will be used. For this application enter
three.
Mode – If more than one reference standard is used, the user must choose between single-phase or
polyphase modes of operation. For this application select polyphase.
Ref Type – This defi nes the reference standard output pull-up value. Select Radian (150 ohm).
Other is 1000 ohm.
Input A** – Select either DUT or REF
SER#1 – Enter a serial number for standard 1, if applicable.
SER#2 – Enter a serial number for standard 2, if applicable.
SER#3 – Enter a serial number for standard 3, if applicable.
* The “
Test By:
” parameter input allows the user to select the type of test – TIME, PULSE,
MANUAL, or ENERGY. If ENERGY is selected, the “
Time:
” parameter input dynamically
changed to “
Energy:
”. This allows a test to be stopped after a pre-set amount of energy has been
accumulated by the D.U.T.
Application 2
Application 2
31
Содержание RM-111
Страница 1: ...Operations Manual Radian Research Inc RM 111 Automated Comparator...
Страница 3: ......
Страница 8: ...RM 111 Operations Manual Product Introduction Product Introduction 7...
Страница 30: ...RM 111 Operations Manual Figure 4 0 Testing a Radian Standard Counting Pulses Application 1 Application 1 29...
Страница 39: ...RM 111 Operations Manual Figure 6 0 Testing a Radian Standard Display Gating Application 3 Application 3 38...