RM-111 Operations Manual
Application 11
Application 11
8. Start the test on the test board. The standard is running on and the test board starts receiving
pulses from the RM-111. After the RM-111 sends the specifi ed number of pulses (based on the Kh
value) to the test board the test will stop.
9. The RM-111’s display shows the Kh value entered earlier in the test and the end count of the
standard(s). The test board itself shows the percent error, which is the percent error of the test
board.