RM-111 Operations Manual
Application 4
Application 4
directly into the keypad and press ENTER. The calculated % registration or % error will appear
at the bottom of the screen. For multiple DUTs enter their respective display reading directly
over the previous display reading. Press ENTER for the calculated result. Use the LEFT and
RIGHT ARROW keys to move the cursor through the display’s digits. The DUT number increases
numerically as multiple entries are made. Repeat this process for all additional DUTs. Press
CANCEL when fi nished to return to the Main Menu.
14. To re-run the test select Re-Run Test. The pulses will be counted again and the new count will
be displayed on the RM-111. Press ENTER to continue.
15. To print the results connect the RM-RPP Remote Portable Printer to the SERIAL PORT on the
RM-111. Turn the RM-RPP to the ON 1 position. Select Print Results. This will start the printing
process. The printout is time/date stamped and includes both the watthour count and the calculated
% Registration or % Error.
16. To save the results select Save Results. The RM-111 will prompt the user for the enabled
parameters in the User Prompts menu and save the test results.
43