Save Run Log File on USB
11.3.2
Power On Logfile
With each system start, the QIAamplifier 96 performs an initial self-test and stores the results in a
power on logfile.
In the power on logfile screen, the following information is listed:
Power-on by date and time
Last power-on by date and time
Last power-down by date and time
A table with a list of error messages by number, date, time, error code and message
Figure 56. QIAamplifier 96 power on log file screen.
11.3.3
Extended self-test log file
During the extended self-test, the QIAamplifier 96 checks several functions and components like
cooler, thermal tracking, heating and cooling rate, refrigeration, gradient, heated lid, and
regulation (see chapter 11.4). Press the corresponding button to see the results for a specific test
summarized in spreadsheet form.