QIAamplifier 96 User Manual 03/2020
96
Note
: The extended self-test will take approx. 30 minutes. The device is blocked while the test is
running and no other program can be started. To stop the extended self-test, press
Cancel Self-Test
.
Cancel Self-Test
If the extended self-test is canceled, either no log file or an incomplete log file will be created.
Figure 61. QIAamplifier 96 extended self-test overview screen.
The results of the extended self-test are summarized in a log file, stored by the device and can be
viewed by the user (see chapter 11.3.3).
11.4
Service info file (SINF)
The service info file is a helpful tool for remote fault diagnosis by QIAGEN Technical Service.
To create a service info file, connect a USB stick to the QIAamplifier 96 and press
Info File to USB
.
Info File to USB
The service info file can be sent through e-mail to QIAGEN Technical Service department. For
contact details, refer to the inside front cover.