QIAamplifier 96 User Manual 03/2020
73
If the heated lid is set to preheating
On
(see chapter 8.5), during the preheating of the heated lid
the word
Preheat
will be displayed in the
Step
field in the table. After the heated lid has reached
the target temperature and the 40s equilibration phase has passed, the first program step starts.
During the program run, the current step is highlighted in yellow in the graphical and spreadsheet
view. In the graphical view, each step is divided into a heating phase and a hold time and both
parts are highlighted separately. Time and temperature increments and ramp rate adjustments
cannot be shown in the graphical view. Instead the symbols
∆
T
,
∆
t
or
!
are displayed (see chapter
8.1).
If it is a gradient program,
Grad
is also displayed in brackets in the
Name
field after the program
name (see Figure 37).
During a gradient step, the temperatures present in each individual column in the sample block can
be checked in the gradient display.
10.2
Pausing a program
To pause a program, press
Pause
on the QIAamplifier 96 run display screen (see Figure
37
).
Pause
Содержание QIAamplifier 96
Страница 21: ...QIAamplifier 96 User Manual 03 2020 21 Figure 1 Voltage selector switch on the underside of the device...
Страница 38: ...QIAamplifier 96 User Manual 03 2020 38 Figure 9 QIAamplifier 96 enter password screen...
Страница 41: ...QIAamplifier 96 User Manual 03 2020 41 Button Block status Color Free Green Running Yellow Pause Blue...
Страница 99: ...QIAamplifier 96 User Manual 03 2020 99 Figure 65 QIAamplifier 96 Cycler Info Screen...
Страница 120: ...QIAamplifier 96 User Manual 03 2020 120 Document Revision History Date Changes March 2020 Initial release...