QIAamplifier 96 User Manual 03/2020
101
12
Quick reference guide
WARNING
Before initial operation, check the correct operating voltage is set on the voltage
selection switch on the underside of the device. Do not open the housing.
Danger of electric shock!
CAUTION
The sample block and the heated lid can reach high temperatures during
operation. Both the sample block and the heated lid can burn you. The samples
are heated up quickly and may come to an explosive boil. Wear goggles when
handling hot samples. Make sure that the lid is closed securely before starting
the program!
NOTICE
The use of oil between samples and sample block is not necessary. If you wish
to use oil, never use silicone oil. Use mineral oil. Ensure that the ventilation slits
are clear.
Important
: Release the lid wheel before opening the lid! Adjust the lid wheel for
every individual run.
Login
After the QIAamplifier 96 has completed the initial self-test, the login screen is displayed:
Press
Login
to log in as an existing user.
Press
Quick Start
for the non-user-specific quick start of programs.
Содержание QIAamplifier 96
Страница 21: ...QIAamplifier 96 User Manual 03 2020 21 Figure 1 Voltage selector switch on the underside of the device...
Страница 38: ...QIAamplifier 96 User Manual 03 2020 38 Figure 9 QIAamplifier 96 enter password screen...
Страница 41: ...QIAamplifier 96 User Manual 03 2020 41 Button Block status Color Free Green Running Yellow Pause Blue...
Страница 99: ...QIAamplifier 96 User Manual 03 2020 99 Figure 65 QIAamplifier 96 Cycler Info Screen...
Страница 120: ...QIAamplifier 96 User Manual 03 2020 120 Document Revision History Date Changes March 2020 Initial release...