![Qiagen EZ2 Connect Скачать руководство пользователя страница 51](http://html2.mh-extra.com/html/qiagen/ez2-connect/ez2-connect_user-manual_3152083051.webp)
EZ2 Connect MDx User Manual 05/2022
51
3.
Check that the voltage rating on the label at the back of the EZ2 matches the voltage
available at the installation site.
4.
Plug the power cord into the instrument power-cord socket.
5.
Plug the power cord into a grounded power outlet.
6.
Do not switch on power to the instrument at this point. Power needs to be OFF for subsequent
installation of USB devices described in the following steps.
WARNING
Damage to electronics
Before powering ON the instrument, make sure that the correct supply voltage
is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the specifications indicated
on the type plate of the instrument.
WARNING
Electrical hazard
Any interruption of the protective conductor (earth/ground lead) inside or
outside the instrument or disconnection of the protective conductor terminal is
likely to make the instrument dangerous.
Intentional interruption is prohibited.
Lethal voltages inside the instrument
When the instrument is connected to line power, terminals may be live and
opening covers or removing parts is likely to expose live parts.
4.3.3
Installation of external barcode scanner
1.
Remove the barcode scanner from the box.
2.
Before powering on the instrument, connect the barcode scanner to any one of the 3 USB
ports located either on the front of the instrument or on the rear of the touchscreen.