EZ2 Connect MDx User Manual 05/2022
140
Figure 144. Select an application screen.
Select and protocol and follow the instructions given on the User Interface to progress through
the various steps to fully set up the protocol, see subsections within the Setting up a protocol
run section (see page 100) for further details.
Note
: Using the EZ2 Connect MDx in Research mode gives the option to skip the initial Load Check.
When skipping the load check, the instrument will not check the worktable setup but immediately
start the protocol run. For the best safety level, QIAGEN recommends to always perform the load
check.
Note
: When reverting the EZ2 Connect MDx to IVD mode it is important to thoroughly clean and
decontaminate the instrument. The following maintenance should be completed:
Section 6.3 Daily maintenance (if not already completed)
Section 6.5 UV decontamination