EZ2 Connect MDx User Manual 05/2022
156
3.
Tap
UV Run
. The date of the last procedure is shown on the screen.
Figure 153. Manual UV Run screen.
4.
Select the number of decontamination cycles. The required decontamination time depends on
the biological material processed on the device.
5.
To begin the procedure, tap
Start
.
6.
If there is a need to abort the UV cycle prior to completion, tap the
Abort
button.
Important
: The UV cycle will not stop immediately upon selecting
Abort
; the system will need
to complete the currently running step of the cycle, this can take 1 or 2 min.
7.
A message (see the figure hereafter) is displayed after execution of the UV Run. Tap
Finish
to
complete the UV Run.
Figure 154. UV run finished screen.