EZ2 Connect MDx User Manual 05/2022
151
Figure 149-B. Installing left side tray.
Figure 149-C. Correct installation of the trays.
10.
Wipe the cartridge and tip racks with 70% ethanol, and then with distilled water.
11.
Wipe the surface of the instrument using a lint-free tissue moistened with 70% ethanol.
Note
: The hood and touchscreen should
ONLY
be cleaned with lint free tissue moistened with
water.