EZ2 Connect MDx User Manual 05/2022
164
9
Technical Specifications
QIAGEN reserves the right to change specifications at any time.
9.1
Operating conditions
Power
100−240 V AC, 50/60 Hz, 1000 VA
Mains supply voltage fluctuations are not to exceed 10%
of nominal supply voltages.
Fuse
AC Inlet: T4A H 250 V
Heater block (temperature fuse): 10A 250V 117C
Overvoltage category
II
Air temperature
18–30°C (64–86°F)
Relative humidity
10–75% RH
Altitude
Up to 2000 m (6500 ft.)
Place of operation
For indoor use only
Pollution level
2
Environmental class
3K21 (IEC 60721-3-3)
3M11 (IEC 60721-3-3)
Average noise level (over 8 hours)
Max. 70 dBA
IP Code (IEC 60529)
IP20
9.2
Transport conditions
Air temperature
−25−60ºC (−13−140ºF) in manufacturer’s package
Note
: If the EZ2 is transported in temperatures below 0ºC
(32ºF), it is recommended to wait 24 hours before
switching on the instrument to allow it to reach the
temperature conditions of the installation environment.
Relative humidity
5–85% RH
Environmental class
2K11 (IEC 60721-3-2)
2M4 (IEC 60721-3-2)
9.3
Storage conditions
Air temperature
5−40°C (41−104°F) in manufacturer's package
Relative humidity
5−85% RH
Environmental class
2K11 (IEC 60721-3-2)
2M4 (IEC 60721-3-2)