EZ2 Connect MDx User Manual 05/2022
157
6.6
Camera calibration
The EZ2 Connect MDx provides an internal camera which allows an inventory scan prior to run
execution.
Note
: The camera calibration should be performed as part of installation procedure, after moving
the instrument, as well as in case of load check issues.
1.
On the touchscreen, tap
Maintenance
.
2.
Tap Camera
LED
.
Figure 155. Camera exposure calibration screen.
3.
Follow the instructions displayed on the screen.
4.
Close the hood.
5.
To begin the procedure, tap
Start
.