9
EDDYSENSOR Operating Manual
Back Module
Front
FI1030BA
EDDYSENSOR from above
Back
The back module of this unit FI
FI
FI
FI
FI 1030BA
1030BA
1030BA
1030BA
1030BA provides a coil connection and
outputs for the analogue defect signal and for the digital defect level. This
unit is supplied in a number of frequency versions:
5 - 10 - 20 - 50 - 100 - 250 kHz.
Coil
FI 1.030BA
Digital
I/O
Analog Signal
0-10V
Output for the defect signal
to a recorder, bar graph or
oscilloscope
Output for the binary defect
signal,
input for end signal
suppression in offline
applications
I/O socket for the coil
Connection of display device
Digital I/O
To connect peripheral devices such as a recorder, bar graph or oscilloscope
on which the signal is graphically displayed, use the socket
Analog Signal 0 - 10V (R
O
= 100 Ohm).
Socket for
output signal
Analog signal
0-10 V
2
3
4
5
1
6
The pins are allocated as follows:
Pin
Use
1
Test signal suppr pole (5 – 30V)
2
Digital ground
3
pole (60V, 100mA), switching duration = min. 200 ms
4
Test signal suppression – pole (5 – 30V)
5
Defect – pole (60V, 100mA)
6
(+24V possible as supply for an external unit, if this is
necessary, Imax = 100mA)
6-pole socket for defect output and
end signal suppression