26
EDDYSENSOR Operating Manual
LAB 1110
connection box
The connection box has two cables to it. These are:
• cable from the pot.free socket on 1050 BA
• mains power cable switched via the EDDYSENSOR
The four outputs are used as follows:
Output
Signal
OUT1
Delayed signal
OUT2
Delayed signal
OUT3
Undelayed signal with a duration of 1 s,
used for the PLC / warning device, e.g. horn
OUT4
System ready signal / coil error
OUT1
OUT2
OUT3
OUT4
Cable from 1050BA
Power cable
to mains
Depending on the modification status of the 1050 module:
In modules with a mod. status up to d, each of the outputs can have a
max. load of 630 mA.
In modules with a mod. status from e, each of the outputs can have a
max. load of 1.25 A
The combined total current drawn may not exceed 2.00 A.
The connection box enables standard peripherals (paint marker, horn etc.)
to be connected directly to the EDDYSENSOR without any rewiring.