Mounting and Installation
Page 42 of 102
Maximum load per loop is 50mA. Short circuit protection at 70mA. Reactivation after 10 seconds. We recommend a maximum
length of 1,000 feet using 22 AWG wire or larger. Twisted, shielded, stranded cable is best for long distances.
If a potential equalization electrode is used, the default wire jumper on XE3 must be removed before connecting the potential
equalization electrode to input XE3:2. If not using a potential equalization electrode, leave the default jumper in place.
FOOTNOTE 1: LM335 is our SGT sensor, pn 1051507
Table 9 pH/ORP-Temp- 4-20mA Input Sensor Connections
Figure 51 pH/ORP=Temp- 4-20mA Input Driver connections
Figure 52 Prominent Industrial
ORP and pH sensors
Table 8 pH/ORP
–
Temperature Connections (Continued next page)