9 Service
■
71
9.3
Calibration,
adjustment
The calibration of the moisture analyser is defined in the configuration
menu (see chapter 3.8 "Weight calibration" and see chapter 5.3.6
"Balance calibration").
9.3.1 Calibrating the balance
Possible types of balance calibration:
• External calibration by means of ICM (Intelligent Calibration Mode)
• External calibration with a freely definable weight
External calibration by means of ICM
Calibration weights in increments of 10g can be used for the moisture
analyser whereby the calibration weights must correspond to the
precision of the instrument.
"SET CALIBRATION MODE EXTERNAL" must be set in the
configuration menu (see chapter 5.3.6 "Balance calibration") for
external calibration by means of ICM.
NOTE
The balance calibration and the temperature calibration can be
interrupted at any time by pressing «
ON/OFF
».
Display
Key
Step
+0,000
9
The balance is in Weighing mode.
+0,000
9
«
T
»
Press until "BALANCE CALIBRA-
TION" apears.
balance CALIBRATION
- - 0000
9
The balance carries out a Zero
measurement "0000 g" is shown
flashing).
- - 50
9
After the zero measurement the dis-
play flashes with the recommended
calibration weight.