7 Ash residue program
■
59
7.1
Ash residue XM 60, XM 60-HR
The XM 60 and XM 60-HR can only store one initial sample weight.
There are no options to set parameters for ash residue determination.
Ash residue:
– Ensure that no moisture analysis is taking place, i.e. that the
instrument is in weighing mode.
Display Key
Step
«
START
»
Keep the key pressed until "ASH
RESIDUE" is displayed. Release
the key.
+
15.000 9
Pour the sampe into the pan
original
0.000 9
+
15.000 9
«
E
»
The initial weight is stored.
original
15.000 9
The initial weight is stored until the ash residue calculation has been done.
End ash residue mode by pressing
«
esc
»
.
+
0.000
9
«
D
»,
«
C
»
If an original weight is stored,
the residual weight is requi-
red when the ash residue is
restarted.
(Select original weight or residu-
al weight.)
RESIDUAL
0.000 9
or
+
0.000
9
orignal
0.000 9
+
9.500
9
Pour the rsediual sample onto
the pan
RESIDUAL
0.000 9
+
63.33
%
«
E
»
Save the residual weight and cal-
culate the ash residue.
ASH RESIDUAL
The ash residue is displayed until any key is pressed to print the ash resi-
due protocol. Than the original and the residual weights are cleared.