6 Determining moisture levels
■
55
End-point condition Digit per time
The drying procedure ends as soon as the reduction in weight, during
the time set, is smaller than the number of digits set. The reduction in
weight must have been greater than the end-point condition at some
point.
In the freely definable mode, the user can set from 1 ... 99 digits in
1 digit increments and from 10...90 seconds in 10 second
increments.
One digit is the smallest change in measurement that can be displayed
by the moisture analyser.
In the XM 60, XM 66:
1 digit = 1 mg
In the XM 60-HR:
1 digit = 0.1 mg
End-point condition % per time
The drying procedure ends as soon as the reduction in weight, during
the time set, is smaller than the % value set. The moist weight
(weight value at the start of the measurement) corresponds to a value
of 100%.
Caution: the reduction in weight must have been greater than the end-
point condition at some point.
In the freely definable mode, the user can set from 0.1% ... 99.9 %
in 0.1 % increments and from 10...90 seconds in 10 second incre-
ments.
ADAPTSTOP
Is a fully automatic stop mode which determines the switch-off time
t [s]
m[Digit]
20 s
60 s
2 Digit (2 mg)
AUTO STOP 2 / 20
AUTO STOP 2 / 60
2 Digit (2 mg)